Home /Guides /Temperature Sensor Calibration: Points, Certificates and Intervals

Temperature Sensor Calibration: Points, Certificates and Intervals

Why calibrate a temperature sensor?

A calibration certificate tells you how much the sensor deviates from a reference at defined points. It does not make the sensor more accurate, but it gives you a documented correction and confidence for quality, safety and maintenance decisions.

Calibration is especially important for:

  • Pharmaceutical and food processes
  • Laboratories and metrology
  • Furnaces and heat treatment
  • Applications with tight accuracy or regulatory requirements

What points should be calibrated?

Choose points that cover the working range, not just one temperature. For a Pt100 used from 0 to 100 °C, common points are 0 °C, 50 °C and 100 °C. For a sensor used around 37 °C, include that point.

For thermocouples, calibration at several points also reveals drift and type errors.

What should the certificate contain?

A useful calibration certificate should include:

  • Sensor identification: model, serial number, range
  • Calibration points and the reference reading
  • Indicated value, error and uncertainty
  • Calibration method and reference standard
  • Date of calibration and recommended next date
  • Laboratory accreditation mark, where applicable

Ask whether the certificate is accredited (for example CNAS) and whether the uncertainty meets your requirement. A certificate with a large uncertainty may not support a tight process tolerance.

Calibration interval

The right interval depends on the sensor, environment and criticality:

  • Laboratory reference sensors: often annual
  • Industrial Pt100 in clean service: 1–2 years
  • Thermocouples in furnaces: 6–12 months, depending on temperature and atmosphere
  • Sensors in critical processes: shorter intervals or before each campaign

A drift history from previous calibrations is the best guide.

In-house vs external calibration

In-house calibration is practical for routine checks with a bath or dry-block calibrator. External calibration is appropriate when you need accredited uncertainty, a wider range or a documented traceability chain.

How to improve calibration results

  • Use a 3- or 4-wire connection for Pt100
  • Keep the sensor and reference at the same depth and in the same medium
  • Allow enough stabilisation time
  • Check the transmitter separately from the sensor
  • Record the as-found and as-left readings

Our high-precision Pt100 sensor can be supplied with a calibration certificate. Use the Pt100 calculator to compare resistance and tolerance values.

Need a calibrated sensor or advice on points and uncertainty? Contact our team.